Ventostat® 8102 Specifications
Measurement Range
0-2000 ppm factory default
Adjustable to 10,000 ppm with UIP Software
Power
18-30 VAC RMS, 50/60 Hz -half-wave rectified 18-42 VDC polarity protected 1.75 VA max avr power 2.75 VA peak power
Operating Conditions
0 – 50oC (32-120°F)
0 – 95% RH, non-condensing
Analog Output (Simultaneous)
0-10 VDC (100 Ohms output impedance)
4-20 mA (RLmax = 500 Ohms)

Accuracy
@60-90°F (15-32°C)
±50 ppm or 5% whichever is greater
(7% for levels over 1500 ppm)
@32-122°F (0-50°C) ±100 ppm or 7% whichever is greater

Relay Output
Normally Open & Normally Closed (wire either way), gold bifurcated, 2A max. @ 24VAC.
Adjustable setpoint (with UIP software kit), factory set at 1000 ppm, 50 ppm hysteresis
Elevation (Pressure) Correction
Add 0.13% of reading per mm Hg decrease from 760 mm Hg (On-board correction set w/UIP software), preset at sea level

Wiring
18-28 AWG stranded copper wire only. 2 wires each for power, analog output, relay

Agency Certification
FCC Part 15 Class B / CE
Digital I/O
RS-232 interface for use with optional PC software and cable in UIP kit 2072
Response Time
0-90% <1 minute
Warranty
18 months parts and labor
Warm-Up Time
@ 25°C <2 minutes

Calibration
5-year calibration guarantee

Storage Temperature
-40 – 70°C
Sample Method
Dual Beam Absorption Infrared™
Diffusion sample method
_______________________________________________________________________________________
©Telaire 1999-2001 - 6489 Calle Real - Goleta California, 93117 - 800-472-6075 e-mail for info
__